Anest. intenziv. Med. 2006;17(4):187-192
Perioperative spinal cord monitoring during spinal surgeryAnaesthesiology - Original Paper
- 1 Klinika anesteziologie, resuscitace a intenzivní medicíny, FN Brno
- 2 Neurologická klinika FN Brno
- 3 Ortopedická klinika FN Brno
Objective:
To compare two methods of perioperative spinal cord monitoring during spinal deformity correction surgery.
Desing:
Observational study.
Setting:
Department of Anaesthesiology and Intensive Care, Department of Orthopaedic Surgery, University Hospital Brno.
Materials and Methods:
Patients admitted to the orthopaedic ICU in 1980-2005 were divided to groups 1-4 of 294, 235, 410 and 116 patients. Patients in groups 1-3 were monitored with the wake-up test in the course of three different anaesthetic techniques. Patients in group 4 were monitored using combined monitoring of somatosensory (SEP) and motor evoked potentials (MEP). We used standard statistical methods with P-value < 0.05 considered significant.
Results:
The wake-up test revealed 3, 3 and 1 cases of paraplegia in groups 1-3 respectively. The paraplegia disappeared after releasing the retraction in all the cases. Three and two cases of clinically relevant paraplegia were diagnosed by SEP and MEP respectively in group 4. None of them lead to a persistent postoperative deficit.
Conclusion:
The authors prefer the methods utilising evoked potentials, because they are a more accurate and continuous indicator of spinal cord ischaemia. The wake up test can be also used.
Keywords: perioperative spinal cord monitoring; surgical correction of spinal deformity; neurological deficit; wake up test; somatosensory and motor evoked potentials
Published: August 1, 2006 Show citation
References
- Cesarová, Z. Anestezie v ortopedii. In Larsen, R. Anestezie. 7. vydání, Praha : Grada Publishing, 2004, 1376 s., ISBN 80-247-0476-5, p. 1290-1293.
- Čundrle, I., Vlach, O. Peroperační monitorování míchy pomocí testu probuzení. Acta chir. orthop. trauma. czech., 1987, 54, 4, p. 347-351.
- Varmužková, J., Vlach, O. Test probuzení - významná součást prevence paraplegie při operační korekci deformit páteře. Acta chir. orthop. trauma. czech., 1982, 49, 5, p. 455-459.
- Němec, M., Bednařík, J., Krbec, M., Mokrá, M. Peroperační monitorace somatosensorických a motorických potenciálů u operací skolioz. Čes. a slov. Neurol. Neurochir., 2006, 69-102, č. 1, s. 45-51.
Go to original source...
- Yeoman, P. M. Influence of induced hypotension and spinal distraction on feline spinal somatosensory evoked potentials. Br. J. Anaesth., 1989, 63, p. 315-320.
Go to original source...
Go to PubMed...
- Machida, M. et al. Dissociation of muscle action potential and spinal somatosensory evoked potential after ischemic damage of spinal cord. Spine, 1988, 13, p. 1119-1124.
Go to original source...
Go to PubMed...
- Andoh, T., Okumura, F. Effects of anestetics on somatosensory and motor evoked potentials in humans. In Stalberg, E., Sharma H. S., Olson, X. Spinal cord monitoring. Springer : Wien, New York 1988.
- Godat, L. et al. Flumazenil et reveil peroperatoire lors de la chirugie de la scoliose. Ann. Fr. Anesth. Reanim., 1990, 9, p. 6-10.
Go to original source...
Go to PubMed...
- Čundrle, I. Řízená hypotenze ve spondylochirurgii. Autoreferát kandidátské dizertační práce, LF MU, 1991, s. 14.
- Němec, M., Bednařík, J., Krbec, M., Leznar, M., Čundrle, I., Mokrá, M. Peroperative monitoring of somatosensory and motor evoked potentials in surgery for scoliosis. Clin. Neurophysiol., 2004, 115, p. 1957-1958.
- Nathan, N. et al. Influence of propofol concentration of multipulse transcranial motor evoked potentials. Br. J. Anaesth., 2003, 91, p. 493-497.
Go to original source...
Go to PubMed...
- Othman, Z. et al. Hypotension-induced loss of intraoperative monitoring during surgical correction of Scheuermann kyphosis: a case report. Spine, 2004, 29, p. 258-265.
Go to original source...
Go to PubMed...
- Padberg, M. et al. Somatosensory and motor evoked potential monitoring without wake up test during idiopathic scoliosis surgery. An excepted standard of care. Spine, 1998, 23, p. 1392-1400.
Go to original source...
Go to PubMed...
- Grundy, B. L. Intraoperative monitoring of sensory-evoked potentials. Anesthesiology, 1983, 87, p. 72-87.
Go to original source...
Go to PubMed...
- Vauzelle, C., Stagnara, P., Jouvinroux, P. Functional monitoring of spinal cord activity during spinal surgery. Clin. Orthop., 1973, 93, p. 173-175.
Go to original source...