Anest. intenziv. Med. 2004;15(4):204-208

Correlation of Cardiac Output Measurement with Transoesophageal Echocardiography and Bolus Termodilution Technique in Patients with Various Degree of Tricuspid RegurgitationArticles

M. Balík, P. Plášil, J. Pažout, M. Otáhal, M. Fric, J. Pachl
Klinika anesteziologie a resuscitace, UK, 3. lékařská fakulta a FNKV, Praha

Objective:
The aim of the article is to suggest a practical approach to the cardiac output (CO) measurement with combineduse of echocardiography and thermodilution in patients with various valve disorders.

Design: Prospective observational study.Setting: 20-bed general intensive care unit in the university hospital.

Material and Methods: 27 mechanically ventilated patients with no significant aortic valve disease were included. Patientswere measured twice using simultaneously transesophageal echocardiography (TEE) and pulmonary artery catheter (PAC)for CO.

Results: Continuous doppler measurements were taken in the left ventricular outflow tract (LVOT) at the level of the aorticvalve.CO was calculated bymultiplying velocity-time integral (VTI) with aortic valve area (AVA) and heart rate. SimultaneousPAC measurements were taken averaging the results of the three 10 cc boluses of iced saline. The difference between themethods expressed as mean ± 2 SD was 0.5 ± 1.1 l/min in the first group of patients (N = 8) with no or the 1st degree tricuspidregurgitration (TR), r = 0.96. A mean difference of 0.8 ± 2.0 l/min was found in the second group (N = 9) with the 2nd degreeTR, r = 0.92. The difference was 1.9 ± 2.3 l/min in the third group (N = 10) with the 3rd degree TR, r = 0.69. Only in the thirdgroup the statistic bias was different from zero (T = 6.83, P < 0.0001).

Conclusions: The presence of significant aortic valve disease on echocardiographic investigation impairs validity of COmeasurement in LVOT. In those cases thermodilution may become the method of choice. On the other hand a high degreeof TR is associated with an underestimation of CO measurement by thermodilution. Echocardiography is more accuratemodality for CO assessment in those patients.

Keywords: tricuspid regurgitation; cardiac output; echocardiography; doppler-ultrasonography; pulmonary arterycatheter; thermodilution

Published: August 1, 2004  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
Balík M, Plášil P, Pažout J, Otáhal M, Fric M, Pachl J. Correlation of Cardiac Output Measurement with Transoesophageal Echocardiography and Bolus Termodilution Technique in Patients with Various Degree of Tricuspid Regurgitation. Anest. intenziv. Med. 2004;15(4):204-208.
Download citation




Anesteziologie a intenzivní medicína

Madam, Sir,
please be aware that the website on which you intend to enter, not the general public because it contains technical information about medicines, including advertisements relating to medicinal products. This information and communication professionals are solely under §2 of the Act n.40/1995 Coll. Is active persons authorized to prescribe or supply (hereinafter expert).
Take note that if you are not an expert, you run the risk of danger to their health or the health of other persons, if you the obtained information improperly understood or interpreted, and especially advertising which may be part of this site, or whether you used it for self-diagnosis or medical treatment, whether in relation to each other in person or in relation to others.

I declare:

  1. that I have met the above instruction
  2. I'm an expert within the meaning of the Act n.40/1995 Coll. the regulation of advertising, as amended, and I am aware of the risks that would be a person other than the expert input to these sites exhibited


No

Yes

If your statement is not true, please be aware
that brings the risk of danger to their health or the health of others.