Anest. intenziv. Med. 2009;20(3):153-157
Blood and tissue glucose level in critically ill patients: Comparison of different methods of measuring interstitial glucose levelsIntensive Care Medicine - Original Paper
- 1 Klinika anesteziologie, resuscitace a intenzivní medicíny Fakultní nemocnice Hradec Králové
- 2 Klinika gerontologická a metabolická, Fakultní nemocnice Hradec Králové
- 3 Centrum pro vývoj a výzkum LF UK a Fakultní nemocnice Hradec Králové
Objective:
The aim of study is to investigate whether the glucose values obtained by miniinvasive techniques of interstitial glucose monitoring (Guardian Real Time (RT) continuous glucose monitoring system and microdialysis) differ substantially from blood glucose readings and whether these methods could be used instead of blood glucose sampling in mechanically ventilated patients with tight glycemic control.
Design:
Local, prospective study.
Setting:
Department of Anesthesiology and Intensive Care, University Hospital.
Materials and methods:
Interstitial glucose concentrations were measured by the Guardian RT system and by the microdialysis of subcutaneous adipose tissue. The microdialysis catheter was constantly perfused with isotonic solution at a flow rate of 0.3 µL/min. Dialysate was collected in hourly fractions. Arterial and venous blood samples were collected in hourly intervals during 24 hours and paired with Guardian RT and microdialysis readings. Data were analyzed using the Bland-Altman method and the correlation coefficient was calculated.
Results:
Correlation coefficient between arterial glycemia and microdialysis interstitial glucose level was r = 0.7044 (P < 0.0001). Correlation coefficient between arterial glycemia and Guardian RT readings was r = 0.6938 (P < 0.0001). Correlation between interstitial glucose level obtained by microdialysis and Guardian RT glucose readings was r = 0.5736 (P < 0.0001).
Conclusion:
Due to obtained results, microdialysis and Guardian Real Time continuous glucose monitoring system could not be considered equivalent to blood glucose measurement and findings do not support their use for tight glycemic control management in mechanically ventilated, critically ill patients.
Keywords: tight glycemic control; interstitial glucose; microdialysis; Guardian Real Time Continuous Glucose Monitoring System
Received: November 24, 2008; Accepted: March 10, 2009; Published: June 1, 2009 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- Van den Berghe, G., Wouters, P., Weekers, F., Verwaest, C., Bruyninck, F., Schetz, M., Vlasselaers, D., Ferdinande, P., Lauwers, P., Bouillon, R. Intensive insulin therapy in the critically ill patiens. N. Engl. J. Med., 2001, 345, p. 1359-1367.
Go to original source...
Go to PubMed...
- Anger, E. K., Szumita, M. P. Barries to glucose control in the intensive care unit. Pharmacotherapy, 2006, 26, 2, p. 214-228.
Go to original source...
Go to PubMed...
- Van den Bergh, G., Schoonheydt, K., Becx, P., Bruynicks, F., Wouters, P. J. Insulin therapy protects the central and peripheral nervous system of intensit care patients. Neurology, 2005, 64, p. 1348-1353.
Go to original source...
Go to PubMed...
- Krinsley, J. S. Effect of an intensive glucose management protocol on the mortality of critically ill adult patiens. Mayo Clin. Proc., 2004, 79, p. 992-1000.
Go to original source...
Go to PubMed...
- Zimmerman, C. R., Mlynarek, M. E., Jordan, J. A., Rajda, C. A., Horst, H. M. An insulin infusion protocol in critically ill cardiothoracic patients. Ann. Pharmacother., 2004, 38, p. 1123-1129.
Go to original source...
Go to PubMed...
- Golberg, P. A., Siegel, M. D., Sherwin, R. S. et al. Implementation of safe and effective insulin infusion protocol in a medical intensive care unit. Diabetes Care, 2004, 32, p. 311-316.
- Critchell, Dana C., Savarese, V., Callahan, A., Aboud, C., Jabbour, S., Marik, P. Accuracy of bedise capillary blood glucose measurements in critically ill patiens. Intensive Care Med., 2007, 33, p. 2079-2084.
Go to original source...
Go to PubMed...
- Ellmerer, M., Haluzik, M., Blaha, J., Kremen, J. Svacina, S., Schaupp, L. Plank, J., Pieber, T. Clinical evaluation of alternative site glucose measurements in patiens after major cardiac surgery. Diabetes Care, 2006, 29, p. 1275-1281.
Go to original source...
Go to PubMed...
- Bahlmann, L., Wagner, K., Heringlake, M., Wirtz, Chr., Futterer, T., Schmucker, P., Klaus, S. Subcutaneous microdialysis for metabolic monitoring in abdominal aortic surgery. J. Clinical. monitoring and computing, 2002, 17, p. 309-312.
Go to original source...
Go to PubMed...
- Aussedat, B., Dupire-Angel, M., Gifford, R., Klein, J. C., Wilson, G. S., Reach, G. Interstitial glucose concentration and glycemia: implications for continuous subcutaneous glucose. Am. J. Physiol. Endocrinol. Metab., 2000, 278, E716-E728.
Go to original source...
Go to PubMed...
- Chetty, V. T., Almulla, A., Odueyungbo, A., Thabane. L. The effect of continuous subcutaneous glucose monitoring (CGMS) versus intermittent whole blood finger-stick glucose monitoring (SBGM) on hemoglobin A1c (HBA1c) levels in Type I diabetic patients: A systematic review. Diabetes Res. Clin. Pract., 2008.
Go to original source...
Go to PubMed...
- Goldberg, P. A., Siegel, M. D., Russell, R. R., Sherwin, R. S., Halickman, J. I., Cooper, D. A., Dziura, J. D., Inzucchi, S. E. Experience with the continuous glucose monitoring system in a medical intensive care unit. Diabetes Technol. Ther., 2004, 6, 3, p. 339-347.
Go to original source...
Go to PubMed...
- Baumeister, F. A., Hack, A., Busch, R. Glucose-monitoring with subcutaneous microdialysis on neonatal diabetes mellitus. Klin. Padiatr., 2006, 218, 4, p. 230-232.
Go to original source...
Go to PubMed...
- Kulkarni, A., Saxena, M., Price, G., O'Leary, M. J. Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients. Intensive Care Med., 2005, 31, 1, p. 142-145.
Go to original source...
Go to PubMed...
- Bedside Glucometry Not Always Accurate in Critically Ill Patients. Mayo Clin. Proc., 2008, 83, p. 400-405.
Go to original source...
Go to PubMed...
- Koschinksy, T. Glucose monitoring: Is the site important? Presented at the 18th International Diabetes Federation Congress, August 25-29, 2003, Paris, France.
- Robinson, L. E., van Soeren, M. H. Insulin resistence and hyperglycemia in critical illness: role of insulin in glycemic control. Am. Assoc. Crit. Care Nurses Clin. Issues, 2004, 15, p. 45-62.
Go to original source...
Go to PubMed...
- Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., Lesho, M. J. Physiological differences between interstitial glucose and blood glucose in human subjects. Diabetes Care, 2003, 26, 8, p. 2405-2409.
Go to original source...
Go to PubMed...
- Brunkhorst, F. M., Engel, Ch., Bloos, F., Meier-Hellmann, A., Ragaller, M. et al. Intensive Insulin Therapy and Pentastarch Resuscitation in Severe Sepsis. N. Engl. J. Med., 2008, 358, p. 125-139.
Go to original source...