Anest. intenziv. Med. 2014;25(4):274-276
Initial experience with the Vivasight ETView® double-lumen fibreoptic tube in thoracic surgeryAnaesthesiology - Original Paper
- Klinika anesteziologie, resuscitace a intenzivní medicíny 2. LF UK ve FN v Motole
Objective:
The double-lumen fibreoptic tube Vivasight ETView® is a new tool that allows lung isolation without the use of a fibrescope. The aim of the study was to evaluate the success and complication rates of its insertion in thoracic surgery procedures.
Design:
Prospective observational study (clinical audit).
Settings:
Department of Anaesthesiology and Intensive Care Medicine, University Hospital.
Material and methods:
We evaluated the success rate, speed of insertion, incidence of malposition and other complications associated with the insertion of a fibreoptic double-lumen tube in 65 patients scheduled for thoracic surgery requiring separation of the lungs.
Results:
The tube was successfully inserted in 64 of 65 patients (98.5%). One patient (1.5%) experienced tube dislocation requiring subsequent repositioning using a flexible bronchoscope. The average speed of insertion was 23.4 (± 4.67) seconds. We did not observe any significant differences in the insertion success rates between experienced anaesthesiologists and trainees.
Conclusion:
The double-lumen Vivasight ETView® fibreoptic cannula is a good tool for the separation of the lungs in thoracic anaesthesia. It is also suitable for educational purposes.
Keywords: thoracic anaesthesia; double-lumen fibreoptic tube; Vivasight ETView®; insertion success rate
Received: January 31, 2014; Accepted: April 2, 2014; Published: August 1, 2014 Show citation
References
- Seymour, A. H., Prasat B., McKenzie, R. J. Audit of double-lumen endobronchial intubation, Br. J. Anaesth., 2004, 93, p. 525-527.
Go to original source...
Go to PubMed...
- Giglio, M., Oreste, D., Oreste, N. Usefulness of ETView TVT endotracheal tube for correct posi-tioning of bronchial blockers in left lobectomy: an easy and safe combination, Minerva Anestesiol., 2009, 75, p. 1-4.
Go to PubMed...
- Huitink, J. M., Koopman, E. M., Bouwman, R. A., Craenen, A., Verwoert, M., Krage, R., Visser, I. E., Erwteman, M., van Groeningen, D., Tinino, R., Schauer, A. Tracheal intubation with a camera embedded in the tube tip (Vivasight(trade)). Anaesthesia., 2013, 68, p. 74-78.
Go to original source...
Go to PubMed...
- Gaitini, L. A., Yanovski, B., Mustafa, S., Hagberg, C. A., Mora, P. C., Vaida S. J. A feasibility study using the VivaSight Single Lumen™ to intubate the trachea through the Fastrach laryn-geal mask airway: a preliminary report of 50 cases. Anesth. Analg., 2013, 116, p. 604-608.
Go to original source...
Go to PubMed...