Anest. intenziv. Med. 2014;25(4):274-276

Initial experience with the Vivasight ETView® double-lumen fibreoptic tube in thoracic surgeryAnaesthesiology - Original Paper

Bicek Vladimír, Krečmerová Martina, Ungermannová Veronika, Vymazal Tomáš*
Klinika anesteziologie, resuscitace a intenzivní medicíny 2. LF UK ve FN v Motole

Objective:
The double-lumen fibreoptic tube Vivasight ETView® is a new tool that allows lung isolation without the use of a fibrescope. The aim of the study was to evaluate the success and complication rates of its insertion in thoracic surgery procedures.

Design:
Prospective observational study (clinical audit).


Settings:
Department of Anaesthesiology and Intensive Care Medicine, University Hospital.

Material and methods:
We evaluated the success rate, speed of insertion, incidence of malposition and other complications associated with the insertion of a fibreoptic double-lumen tube in 65 patients scheduled for thoracic surgery requiring separation of the lungs.

Results:
The tube was successfully inserted in 64 of 65 patients (98.5%). One patient (1.5%) experienced tube dislocation requiring subsequent repositioning using a flexible bronchoscope. The average speed of insertion was 23.4 (± 4.67) seconds. We did not observe any significant differences in the insertion success rates between experienced anaesthesiologists and trainees.

Conclusion:
The double-lumen Vivasight ETView® fibreoptic cannula is a good tool for the separation of the lungs in thoracic anaesthesia. It is also suitable for educational purposes.

Keywords: thoracic anaesthesia; double-lumen fibreoptic tube; Vivasight ETView®; insertion success rate

Received: January 31, 2014; Accepted: April 2, 2014; Published: August 1, 2014  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
Bicek V, Krečmerová M, Ungermannová V, Vymazal T. Initial experience with the Vivasight ETView® double-lumen fibreoptic tube in thoracic surgery. Anest. intenziv. Med. 2014;25(4):274-276.
Download citation

References

  1. Seymour, A. H., Prasat B., McKenzie, R. J. Audit of double-lumen endobronchial intubation, Br. J. Anaesth., 2004, 93, p. 525-527. Go to original source... Go to PubMed...
  2. Giglio, M., Oreste, D., Oreste, N. Usefulness of ETView TVT endotracheal tube for correct posi-tioning of bronchial blockers in left lobectomy: an easy and safe combination, Minerva Anestesiol., 2009, 75, p. 1-4. Go to PubMed...
  3. Huitink, J. M., Koopman, E. M., Bouwman, R. A., Craenen, A., Verwoert, M., Krage, R., Visser, I. E., Erwteman, M., van Groeningen, D., Tinino, R., Schauer, A. Tracheal intubation with a camera embedded in the tube tip (Vivasight(trade)). Anaesthesia., 2013, 68, p. 74-78. Go to original source... Go to PubMed...
  4. Gaitini, L. A., Yanovski, B., Mustafa, S., Hagberg, C. A., Mora, P. C., Vaida S. J. A feasibility study using the VivaSight Single Lumen™ to intubate the trachea through the Fastrach laryn-geal mask airway: a preliminary report of 50 cases. Anesth. Analg., 2013, 116, p. 604-608. Go to original source... Go to PubMed...




Anesteziologie a intenzivní medicína

Madam, Sir,
please be aware that the website on which you intend to enter, not the general public because it contains technical information about medicines, including advertisements relating to medicinal products. This information and communication professionals are solely under §2 of the Act n.40/1995 Coll. Is active persons authorized to prescribe or supply (hereinafter expert).
Take note that if you are not an expert, you run the risk of danger to their health or the health of other persons, if you the obtained information improperly understood or interpreted, and especially advertising which may be part of this site, or whether you used it for self-diagnosis or medical treatment, whether in relation to each other in person or in relation to others.

I declare:

  1. that I have met the above instruction
  2. I'm an expert within the meaning of the Act n.40/1995 Coll. the regulation of advertising, as amended, and I am aware of the risks that would be a person other than the expert input to these sites exhibited


No

Yes

If your statement is not true, please be aware
that brings the risk of danger to their health or the health of others.